Product core introduction
Product NameHAST Unsaturated High Pressure Accelerated Aging Chamber
Sample limitThis test equipment is prohibited:
Testing or storage of flammable, explosive, volatile materials sample .
Testing or storage of corrosive substance samples.
Testing or storage of strong electromagnetic emission source samples.
Performance
Test environmental conditions
Test methods
The environmental temperature is +25℃, the relative humidity is ≤85%, no test sample in the test chamber
GB/T 5170.2-1996 temperature test equipment.
Temperature range105℃ → +132℃. Or 105℃ → +142.9℃.
Temperature fluctuation≤±0.5℃.
Temperature uniformity≤±2.0℃.
Humidity range65% ~ 100 %R.H .
Humidity fluctuation≤±2.5 %R.H.
Humidity uniformity≤±3.0% R.H.
Pressure range0.2~2㎏/㎝2 (0.05~0.196 MPa)
Heating timeNormal temperature→ + 132℃ ,about 50 min
Pressure rising timeAtmospheric pressure→ + 2㎏/㎝2 , about 45 min
External air supply pressure: about 5 min
Load situationMagnetic material
Bias terminalNo bias terminal (specify when ordering)
Satisfied standard
IEC60068-2-66; JESD22-A102-C; JESD22-A110-B;JESD22-A118;
GB/T 2423.40-1997
IEC 60068-2-66,an environmental testing standard of the IEC (International Electrotechnical Commission).